The Fluke Calibration 3130 is built into a rugged case with internal storage for power supply, test leads, and fittings. In addition, it includes a rechargeable, NiMH battery, allowing for approximately 50 hours of operation. When using the internal pump, the battery is sufficient to provide up to 100 calibration cycles to 300 psi. The battery is recharged by simply plugging the unit in using the included universal mains adapter.
Specifications
Environmental
Operating temperature
-10 °C to +50 °C
Storage temperature
-20 °C to +60 °C
Power requirements
12 V dc (Universal ac adapter/charger supplied)
Battery
Internal 3800 mAh advanced NiMH pack
Operating time
On full charge
Approx 50 hours (Measure only or external air; no pump)
Using internal pump
Approx 100 calibration cycles to 300 psi
Physical
Dimensions
15.25 in L x 12 in W x 7 in D
Weight
~7 kg (15 lb)
EMI/RFI conformance
EN61326:2006 Annex A
Connectors/ports
1/8 in NPT (External supply port and test port)
Included accessories
Manual, NIST-traceable certificate, test leads, universal ac adapter/charger
Ranges
Pressure (internal pump)
-80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar)
Pressure (external air)
0 to 2 MPa (0 to 300 psi, 0 to 20 bar)
mA
0 to 24.000 mA
Volts
0 to 30.000 V dc
Engineering units
psi, bar, mbar, kPa, MPa, kgf/cm2, mmH2O @ 4 °C, mmH2O @ 20 °C, cmH2O @ 4 °C, cmH2O @ 20 °C, inH2O @ 4 °C, inH2O @ 20 °C, inH2O @ 60 °F, mmHg @ 0 °C, inHg @ 0 °C
Instrumental measurement uncertainty
Pressure
0.025 % of reading ± 0.01 % FS
mA
0.015 % of reading ± 0.002 mA
Volts
0.015 % of reading ± 0.002 V
Temperature effect (all functions)
No effect on accuracy on all functions from 15 °C to 35 °C
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C