Ametek JOFRA RTC-250: High accuracy, excellent stability 0.01C, wide temperature range. Range: 28 to 250°C
Advantages of the combined liquid bath/dry-block calibrator
- • Automatic calibration of as many as 24 sensors at a time
- • For customers, who only want to use liquid baths
- • For calibration of odd sizes and shapes of sensors including sanitary sensors
WET = no need for inserts, that fit the sensors
DRY = more space for calibration of special sensors
- • Industries who need to calibrate many sensors at a time or short sensors can benefit from the big well
- • JOFRACAL software and RTC B-models can handle on-line calibration and documentation of multiple sensors calibrated simultaneously
FUNCTIONAL SPECIFICATIONS
Temperature range
RTC-158
@ ambient temp. 0°C/32°F.........-37 to 155°C/-35 to 311°F
@ ambient temp. 23°C/73°F.........-22 to 155°C/-8 to 311°F
@ ambient temp. 40°C/104°F........ -9 to 155°C/16 to 311°F
RTC-250
@ ambient temp. 0°C/32°F..............5 to 250°C/41 to 482°F
@ ambient temp. 23°C/73°F..........28 to 250°C/82 to 482°F
@ ambient temp. 40°C/104°F......45 to 250°C/113 to 482°F
Accuracy (model B & C) with external STS ref. sensor
RTC-158 B & C.......................................... ±0.04°C/±0.07°F
RTC-250 B & C.......................................... ±0.07°C/±0.13°F
12-month period. Relative to reference standard. Specs by use of
the external STS-200 reference sensor. Excl. sensor drift.
Accuracy with internal reference sensor
RTC-158 A, B & C...................................... ±0.18°C/±0.32°F
RTC-250 A, B & C...................................... ±0.28°C/±0.50°F
Stability
RTC-158................................................... ±0.01°C/±0.018°F
RTC-250................................................... ±0.02°C/±0.036°F
Measured after the stability indicator has been on for 15 minutes.
Measuring time is 30 minutes.
Radial homogeneity (difference between holes)
RTC-158 @ -22°C/-8°F, Block........................0.03°C/0.05°F
RTC-158 @ 155°C/311°F, Block.....................0.05°C/0.09°F
RTC-158 @ range, Bath................................0.015°C/0.03°F
RTC-250 @ range, Block.................................0.05°C/0.09°F
RTC-250 @ range, Bath................................0.015°C/0.03°F
Resolution (user-selectable)
All temperatures .......................1° or 0.1° or 0.01° or 0.001˚
Heating time
RTC-158 -22 to 23°C/-8 to 73°F.................... 9 minutes
23 to 100°C/73 to 212°F............... 23 minutes
100 to 155°C/212 to 311°F........... 28 minutes
RTC-250 28 to 100°C/82 to 212°F................. 3 minutes
50 to 100°C/122 to 212°F............... 2 minutes
100 to 250°C/212 to 482°F............. 9 minutes
Cooling time
RTC-158 155 to 100°C/311 to 212°F............. 9 minutes
100 to 23°C/212 to 73°F............... 24 minutes
23 to 0°C/73 to 32°F..................... 15 minutes
0 to -15°C/32 to 5°F..................... 21 minutes
RTC-250 250 to 100°C/482 to 212°F........... 27 minutes
100 to 50°C/212 to 122°F............. 27 minutes
50 to 28°C/122 to 82°F................. 28 minutes
Time to stability (approx.)
RTC-158.............................................................. 15 minutes
RTC-250.............................................................. 15 minutes
Immersion depth
RTC-158/250 incl. insulation plug................. 180 mm/7.1 in
RTC-158/250 bath version............................ 150 mm/5.9 in
PHYSICAL SPECIFICATIONS
Weight and instrument size (L x W x H)
RTC-158........................................................... 11 kg/24.3 lb
RTC-250.........................................................9.9 kg/ 21.8 lb
RTC-158/250..... 366 x 171 x 363 mm / 14.4 x 6.7 x 14.3 in
Shipping (without carrying case)
RTC-158........................................................... 17 kg/37.5 lb
RTC-250........................................................... 16 kg/35.3 lb
Size.......................580 x 250 x 500 mm/22.8 x 9.8 x 19.7 in
Shipping (including optional carrying case)
RTC-158........................................................... 28 kg/61.7 lb
RTC-250........................................................... 27 kg/59.6 lb
Size.....................550 x 430 x 660 mm/21.7 x 16.9 x 26.0 in
INSERTS
Insert dimensions
RTC-158/250 outer diameter........................63.5 mm/2.5 in
RTC-158/250 length.......................................160 mm/6.3 in
Weight of non-drilled insert (approx.)
RTC-158/250................................................. 1200 g/42.3 oz
Alloy
RTC-158/250................................... Special aluminium alloy
Use of other inserts may reduce performance of the calibrator.
To get the best results out of the calibrator, the
insert dimensions, tolerance and material is critical. We
highly advise using JOFRA inserts, as they guarantee