Simpson 260-6XLM & 260-6XLPM Low Power VOM
Manufacturer: Simpson
Part Number: 122
Product Options
* denotes required field
Simpson 260-6 Model Options*
NIST Traceable Calibration Certificate*
|
|
Simpson 260-6XLM & 260-6XLPM Low Power VOM
The 260-6XLM has extra low power ohms ranges. It also offers all of the advantages of the standard 260 family, including a resettable overload protection circuit on the 260-6XLPM. Both versions come with batteries, manual and test leads with alligator clips.
Features:
- 2 extra "low power" ohms ranges: Rx1 and Rx10
- Conducts in-circuit test without risk to components or causing digital circuits to switch states
- Durable ABS plastic case provides impact resistance and drop protection
- Large, tri-colored dial with mirrored scale to prevent parallax error
- Instant trend information for quick positive "Yes/No" checks
- Mirrored Scale for Parallax Reduction
- Resettable overload circuit protection relay
Technical Specifications:
DC Voltage:
|
Ranges |
Accuracy |
Sensitivity |
|
250 mV, 1V, 2.5V, 10V, 25V, 100V, 250V, 500V, 1000V |
2% of full scale |
20 KOhm per volt |
AC Voltage:
|
Ranges |
Accuracy |
Sensitivity |
|
2.5V, 10V, 25V, 100V, 250V, 500V, 1000V |
3% of full scale |
5 KOhm per volt |
Decibel:
|
5 Ranges: |
-20 to 49 dB |
|
Reference: |
0dB= 1mW into 600 Ohm |
AC Current: Use up to 250 AC amperes with AC Amp-Clamp Adapter Accessory 00541
DC Current:
|
Ranges |
Accuracy |
Voltage Drop: (Nominal) |
|
50uA, 0.5mA, 5mA, 50mA, 500mA, 5A |
1% of full scale 50 uA 2.0% full scale, all other |
250mV on 50uA/0.5mA/5A 252mV on 5mA/50mA 400mV on 500mA |
Standard Power Resistance:
|
Ranges |
Accuracy |
Maximum Indication |
Open Circuit Voltage |
Short Circuit Current |
|
Rx1 |
2.5 Deg of arc on Rx1 2.0 Deg of arc on all others |
0-2 KOhms |
1.5V |
250mA |
|
Rx100 |
0-200 KOhms |
1.5V |
2.5mA |
|
Rx1K |
0-2 MOhms |
1.5V |
250uA |
|
Rx10K |
0-20 MOhms
|
9.0v |
125uA |
Low Power Resistance:
|
Ranges |
Accuracy |
Maximum Indication |
Open Circuit Voltage |
Short Circuit Voltage |
|
Rx1 |
2.5 Deg of arc on Rx1 |
0-1 KOhms |
100mV |
4.9mA |
|
Rx10 |
2.0 Deg of arc on Rx10 |
0-10 KOhms |
100mV |
490uA |
General Specifications:
Batteries: 1.5V D, 9V (NEDA No. 13F, 1604A) Fuse: F1 1A, 250V, type 3AG, quick acting Littelfuse® Type 312001, F2 2A, 600V Littelfuse® Type BLS or 2A 600V, Bussman® Type BBS Size: 5 1/4" x 7" x 3 1/8" (13.34 x 17.78 s 7.94 cm) Weight: 2 1/4 lbs. (1.02 Kg) Rated Circuit-To-Ground Voltage: 1000V AC/DC Max
|
|


|
|